Cryolab Publishes Guide to Choosing Cryoboxes and Cryogenic Freezer Boxes

eTradeWire News/10853552
Technical guidance on footprint, grid format, height and material for laboratories planning cryogenic sample storage, including why boxes belong in vapour phase rather than liquid nitrogen.

CHICHESTER, U.K. - eTradeWire -- Cryolab, the cryogenic equipment and consumables supplier based in Chichester, has published a technical guide to selecting cryoboxes and cryogenic freezer boxes for laboratory sample storage.

The guide addresses a choice that is usually made by habit rather than specification. The box determines how many vials fit in a rack, how long a retrieval takes at the open neck of a storage vessel, and whether the inventory record still matches the contents a year later.

Three decisions are covered in detail. Footprint and height: almost all laboratory cryoboxes share a nominal 133 x 133 mm footprint, the dimension that mechanical freezer racks, drawer systems and vapour phase shelving are built around, with external heights of 25 mm for 0.5 mL microtubes, 50 to 52 mm for 1.0 to 2.0 mL vials, 75 mm for 4 to 5 mL vials and 95 to 100 mm for tall formats.

More on eTradeWire News
Grid format: within the same footprint, dividers give 100 positions at 10 x 10, 81 at 9 x 9, 36 at 6 x 6 and 25 at 5 x 5, which across a rack of five boxes works out at 500, 405, 180 or 125 vials. The guide notes that a 100 position box holds 23 per cent more than an 81 position box but is slower to work in with gloved hands, and recommends leaving one box position per rack free so that a store can be reorganised or absorb a transfer out of a failed vessel.

Material: polycarbonate is commonly specified to around -86 degrees C and becomes brittle as temperature falls, while coated board is commonly rated to -196 degrees C but absorbs moisture and can stick in the rack.

The guide is firm on one point. Cryoboxes should not be submerged in liquid nitrogen. A submerged box fills with liquid, and a vial with an imperfect seal takes it inside. One litre of liquid nitrogen becomes roughly 683 litres of gas at 1.013 bar and 15 degrees C, so a vial that has taken on liquid can rupture when warmed. Boxes belong in a mechanical freezer at -80 degrees C or in a vapour phase rack above the liquid.

More on eTradeWire News
Cryolab supplies cryoboxes and cryogenic freezer boxes to order in the standard footprint across the common grid formats and heights, alongside cryotubes, cryovials and a wider range of laboratory consumables and cryogenic accessories. The company was established in 2000, holds ISO 9001:2015 certification and works with IVF clinics, hospitals, biobanks and research laboratories internationally.

The guide is available at https://cryolab.co.uk/cryoboxes-and-cryogenic-freezer-boxes-choosing-grid-height-and-material/